DSpace Repository

Scleral Perforation following Transscleral Diode Laser Cycloablation:

Show simple item record

dc.contributor.author Afshan Ali
dc.date.accessioned 2026-08-13T05:44:22Z
dc.date.available 2026-08-13T05:44:22Z
dc.date.issued 2025-04-01
dc.identifier.issn 2220-7562
dc.identifier.uri http://hdl.handle.net/123456789/21564
dc.description.abstract Transscleral Diode laser cycloablation (DLCA) is one of the cycloablative procedures usually used for treating cases of refractory glaucoma. We report here a case of 80 years old man, who underwent transscleral diode laser cycloablation in his left painful blind eye. The main aim of this treatment was to lower down the intraocular pressure so that his pain could be relieved. On regular follow up visit, just two weeks after the procedure, he presented with scleral perforation on inferior 180 degrees of eye exactly where laser was applied. His intraocular pressure was 12mmHg. Patient had to undergo tectonic and conjunctival grafting on emergency basis to save the integrity of the globe and to avoid phthisis, pain and infection. His post op intraocular pressure (IOP) was 10mmHg with graft in place. This is a very rare complication and to the authors knowledge, this is first reported case of scleral perforation following transscleral diode laser cycloablation in Pakistan en_US
dc.description.sponsorship BAHRIA UNIVERSITY en_US
dc.language.iso en en_US
dc.publisher Bahria University Health Sciences Campus Karachi en_US
dc.relation.ispartofseries 15;02
dc.subject Glaucoma, Intraocular pressure, diode laser en_US
dc.title Scleral Perforation following Transscleral Diode Laser Cycloablation: en_US
dc.type Article en_US


Files in this item

This item appears in the following Collection(s)

Show simple item record

Search DSpace


Advanced Search

Browse

My Account